| STEP |
WHO |
WHEN |
INFO TO USE |
WHAT TO DO |
RESULTS |
| 1 |
Any
team member |
Before using any precision measuring instrument |
The
calibration tag
on the instrument and the
Measuring Instrument Verification Log |
At
the start of the each shift:
1.Verify1 that each
instrument and/or standard to be used is tagged with a
current Trinity Forge "Calibration" sticker.
2. Verify
each quality related measuring instrument that will be
used during a shift. Record the verification and all other pertinent
information on the Measuring Instrument
Verification Log. IF THE INSTRUMENT and/or STANDARD is DAMAGED, HAS
AN EXPIRED CALIBRATION STICKER, FAILS TO ZERO, OR FAILS VERIFICATION
AGAINST A CURRENT CALIBRATED STANDARD, SUBMIT IT IMMEDIATELY TO THE
QC SUPERVISOR or METROLOGIST2---DO NOT USE!!
|
Only calibrated instruments are used for measurements that will affect
the quality of manufactured parts |
| 2 |
QC
Supervisor or Metrologist |
Instrument(s) needs to be repaired or adjusted |
Verbal and/or visual information from any team member about the
instrument(s) condition |
Submit the instrument(s) to Metrology. Initiate recheck
per
TOP-RECHK |
Defective instrument(s) is submitted to Metrology. Recheck procedure is
initiated |
| 3 |
Any
team member (typically QC
Supervisor) |
Having any precision measuring instrument needing to be placed into the
calibration system |
Personal knowledge |
Submit instrument to Metrology for an acceptance
review. |
The instrument is submitted to Metrology for review |
| 4 |
Metrologist |
Upon receiving a precision instrument for entry or update into calibration
control system |
Request from submitter |
Calibrate the instrument per step 6. |
Perform the precision instrument calibration4
process outlined in step 6 |
| 5 |
Metrologist |
As
instruments become due for periodic calibration
or are submitted for calibration due to questionable performance |
Calibration log |
Collect all instruments that are due or that have been
turned in by users for calibration. Identify in
calibration log any instruments that have been lost,
discarded or taken out of service using
Calibration Control
System. |
Instruments are accumulated and ready for calibration or
disposition recorded in calibration log |
| 6 |
Metrologist |
As instruments are submitted for calibration and as due dates make it
necessary |
Calibration log; precision instrument manufacturer info
file |
Calibrate the instrument to standards traceable to NIST and record the
readings in the
Calibration Control System.
For Quality Control Inspectors and others required to meet the
SNT-TC-1A-8.2 Visual Acuity Standard, refer the individual to Human
Resources for the necessary eye examination. Magnaflux Inspectors are
required to have training by a certified Magnaflux Inspector to maintain a
Magnaflux Inspector Level II rating. Furnace temperature controls are
calibrated per
TOP-BRKDN. Perform any instrument
adjustment as needed to bring the
instrument within the
acceptable limits for proper calibration. For an instrument
that is the property of Trinity Forge, Inc. requiring advanced or
extensive calibration,
adjustment or repair, submit the instrument
to an approved calibration vender capable of performing
the required calibration,
adjustment or repair per the ANSI/NCSL Z540-1-1994 standard and/or
ASME-B-89 standard for software involved calibration,
adjustment or repair. Any personal instruments
requiring advanced or extensive calibration,
adjustment or repair will be returned to the owner for disposition. ALL transactions will
be noted in the Calibration Log. Cite applicable
Federal specifications and tolerances along with required traceability to
NIST. Require a report on the condition of the instrument
as received by the calibration vender. |
A pass or fail determination is made on the instrument. The Calibration
Control System updated |
| 7 |
Metrologist |
For any instrument failing calibration (including vender calibrations) that
cannot be economically repaired |
Results of calibration |
Tag
the defective instrument with a red REJECT label and make the appropriate
entry in the Calibration Log. Return personal or
customer instruments to the owner. Properly dispose of
Trinity Forge defective instruments. Notify the supervisor(s) of affected area. |
Defective instrument is removed from premises |
| 8 |
Metrologist |
For any instrument failing calibration (including vendor calibrations) that
can be economically repaired |
Results of Calibration |
Notify the QC Supervisor. Repair instrument in-house,
if possible or send to an approved calibration vendor per step 6 and
initiate re-checking per TOP-RECHK. In the case of Visual Acuity failure,
after corrective action is taken, refer to step 6. |
Defective instrument rechecking procedure initiated |
| 9 |
Metrologist |
When an instrument passes calibration |
Results of calibration |
Decide on the next calibration date based on type, condition, environment,
use of the instrument and recommended calibration
frequency5. Update the calibration log and tag the
instrument with a calibration label showing
identification, calibration date, next calibration date and the
calibrator’s stamp. Apply any safeguard to protect the instrument
from adjustments that would invalidate the calibration.
Return the instrument to user with any recommendations.
|
The
calibrated instrument is tagged, logged, safeguarded
from adjustments, and returned |
| 10 |
Metrologist |
When a calibration date needs to be extended |
Calibration log or own judgment |
Calibration due dates may be extended by no more than one-half of its
normal calibration period under circumstances deemed appropriate by the
Metrologist. Note the extension and circumstances in
calibration log. |
The calibration date is extended for the calibrated instrument and the
calibration log annotated |
| 11 |
Metrologist |
When instruments are not in use for extended periods or are missing |
Personal knowledge |
For
an instrument that is not in use it may be tagged as
“IA” (inactive), appropriately stored and noted in the calibration log.
Tools that are missing will be entered into the calibration log as such. A
“Tool Status” e-mail is to be sent to the appropriate supervisor(s). |
All tools tagged inactive or are missing are appropriately disposed. The
calibration log is updated and the affected supervisor(s) are emailed |
| 12 |
Metrologist, Engineer, or Supervisor |
When an inactive ("IA") precision measuring instrument is needed |
Personal knowledge and the calibration tag on instrument |
Go to step 4 unless the Metrologist is not available. In the event that the
Metrologist is not available, and a tool is needed in an emergency, a
supervisor may temporarily activate the tool by doing a calibration per step
6. An email must be sent immediately to Metrologist. When the Metrologist
becomes available, the tool will be submitted for proper calibration and
entry into the calibration log. |
Metrologist calibrates and updates calibration control system |
1
An instrument is verified by taking measurements at the lower 25% AND upper
75% (rounded to the nearest whole increment) of the measuring instrument's
measurement range. After taking the above maximum measure, the measuring
instrument must be returned to zero (0) point and the reading observed. All prescribed tolerances must be observed. Deficiencies including tip wear
that may cause the instrument to measure incorrectly must be monitored. 2 Metrologist" refers to the Metrologist or a designee. Typically the
Metrologist Assistant or the Machining Department QC Lead man.
3
Includes, but is not restricted to any existing calibrated and logged
precision measuring device (caliper, CMM, v-block, micrometer, scale,
fixture, gage etc), any precision measuring device not previously entered into
the calibration system, any previously rejected precision measuring device
that has been repaired, or any individual required by TOPS (QC Inspectors,
Magnaflux inspectors, etc) to be entered into the calibration system.
4
Metrologist calibrations are per ANSI/NCSL Z540-1-1994.Readings
must be taken at the beginning point, mid point, end point and back to
beginning point of each instrument with no greater than a three (3) inch
increment between calibration points unless otherwise specified by the
Metrologist. On calipers, tip-wear must be
monitored by taking readings well into the jaws as well as near the
beginning of the jaws. Tolerances for calipers and similar instruments are
+/- .003 inches in the Forging Department and +/- .002 inches in the
Machining Department. Tolerances for micrometers are to +/- .002 inches in
Forging Department and +/-.0005 inches (with 10th graduations) or +/- .001
inches (without 10th graduations) in the Machine Department. Machine
Department Gage Pins verification is +/-.0003. Exceptions, other
instruments must be accurate to within 5% of the dimension tolerance
interval of the dimension being measured. Exceptions must come from
the QC
Supervisor.
5
Calibration frequency is 12 months for CMM, Surface Tables, Master
Blocks, CNC Machines, Thread Gages, and 1-5 years (based on usage) on
Standard Blocks. Visual Acuity examinations will be
given within the first ninety (90) days of hire date regardless of temporary
or full time employment status. This testing is required for QC Inspectors, QC Leadman,
Post Forge Workers that work on the inspection table, or anyone doing a visual
inspection of a part as part of final review. This testing will be
administered annually and the employee will be identified with an “Eyeballs”
sticker and calibration
date on their hard hat. Magnaflux Level II every 3 years. Calibration
frequency on all other instruments is 6 months unless otherwise specified by
the Metrologist.
6 Calibration Control System
includes but is not restricted to Metrology, Quality Control, the
Calibration Log, the Verification Log, and Gage Manager.
7
Current
Trinity Forge calibration stickers include but are not restricted to
Trinity Forge calibration stickers and third party calibration stickers
bearing a Metrologist signature or stamp.
8
Flight Safety Requirements All measurement and test
equipment used to inspect Flight Safety Parts is discriminate to within ten
percent (10%) of the total tolerance spread for the feature being inspected.
For total tolerance spreads of less than .001 inch, measurement equipment is
discriminate to within twenty percent (20%) of the spread. Equipment less
discriminate is used with the specific approval of the customer.