| STEP |
WHO
|
WHEN
|
INFO TO USE
|
WHAT TO DO
|
RESULTS
|
| 1 |
Any team
member |
Before using
any precision measuring
instrument |
The calibration
tag on the instrument and the Measuring Instrument Verification
Log |
At the start of
the each shift:
1. Verify1 that each
instrument and/or standard to be used is tagged
with a current Trinity Forge
"Calibration"
sticker.
2. Verify each
quality related measuring
instrument2 that will be used
during a shift;record the verification and all
other pertinent information on the Measuring
Instrument Verification Log; IF THE INSTRUMENT
and/or STANDARD is DAMAGED, HAS AN EXPIRED
CALIBRATION STICKER, FAILS TO ZERO, OR FAILS
VERIFICATION AGAINST A CURRENT CALIBRATED STANDARD,
SUBMIT IT IMMEDIATELY TO THE QC SUPERVISOR
or CALIBRATIONS
CLERK--DO NOT USE!!
|
Only calibrated
instruments are used for measurements that will
affect the quality of manufactured
parts |
| 2 |
QC Supervisor
or Calibrations Clerk |
Instrument(s)
needs to be repaired or
adjusted |
Verbal and/or
visual information from any team member about the
instrument(s) condition |
Submit the
instrument(s) to Metrology. Initiate recheck per
TOP-RECHK. |
Defective
instrument(s) is submitted to Metrology. Recheck
procedure is initiated |
| 3 |
Any team member
(typically QC Supervisor) |
Having any
precision measuring instrument needing to be placed
into the calibration system |
Personal
knowledge |
Submit instrument
to the Calibrations Clerk for
an acceptance review. |
The instrument is
submitted to the Calibrations
Clerk for review |
| 4 |
Calibrations
Clerk
|
Upon receiving a
precision instrument for entry or update into
calibration control system |
Request from
submitter |
Calibrate the
instrument per step 6. |
Perform the
precision instrument calibration3
process outlined in step 6 |
| 5 |
Calibrations
Clerk |
As instruments
become due for periodic calibration or are submitted
for calibration due to questionable
performance |
Calibration
log |
Collect all
instruments that are due or that have been turned in
by users for calibration; identify in calibration log
any instruments that have been lost, discarded or
taken out of service using Calibration
Control System |
Instruments are
accumulated and ready for calibration or disposition
recorded in calibration log |
| 6 |
Calibrations
Clerk |
As instruments
are submitted for calibration and as due dates make
it necessary |
Calibration log;
precision instrument manufacturer info
file |
Calibrate the
instrument to standards traceable to NIST and
record the readings in the Calibration Control
System; for
Quality Control Inspectors and others required to
meet the SNT-TC-1A-8.2 Visual Acuity Standard,
refer the individual to Human Resources for the
necessary eye examination; magnaflux Inspectors are
required to have training by a certified Magnaflux
Inspector to maintain a Magnaflux Inspector Level
II rating; furnace temperature controls are
calibrated per
TOP-BRKDN; perform any
instrument adjustment as needed to bring the
instrument within the acceptable limits for proper
calibration; for an instrument that is the property
of Trinity Forge, Inc. requiring advanced or
extensive calibration, adjustment or repair, submit
the instrument to an approved calibration vender
capable of performing the required calibration,
adjustment or repair per the ANSI/NCSL Z540-1-1994
standard and/or ASME-B-893
standard for software involved calibration,
adjustment or repair;any personal instruments
requiring advanced or extensive calibration,
adjustment or repair will be returned to the owner
for disposition; all transactions will be noted in
the Calibration Log; cite applicable Federal
specifications and tolerances along with required
traceability to NIST; require a report on the
condition of the instrument as
received by the calibration
vender.
|
A pass or fail
determination is made on the instrument. The
Calibration Control System
updated |
| 7 |
Calibrations
Clerk |
For any
instrument failing calibration (including vender
calibrations) that cannot be economically
repaired |
Results of
calibration |
Tag the defective
instrument with a red REJECT label and make the
appropriate entry in the Calibration Log; return
personal or customer instruments to the owner;
properly dispose of Trinity Forge defective
instruments. Notify the supervisor(s) of affected
area. |
Defective
instrument is removed from
premises |
| 8 |
Calibrations
Clerk |
For any
instrument failing calibration (including vendor
calibrations) that can be economically
repaired |
Results of
Calibration |
Notify the QC
Supervisor. Repair instrument in-house, if possible
or send to an approved calibration vendor per step 6
and initiate re-checking per TOP-RECHK; in the case
of Visual Acuity failure, after corrective action is
taken, refer to step 6. |
Defective
instrument rechecking procedure
initiated |
| 9 |
Calibrations
Clerk |
When an
instrument passes
calibration |
Results of
calibration |
Decide on the
next calibration date based on type, condition,
environment, use of the instrument and recommended
calibration frequency4; update the
calibration log and tag the instrument with a
calibration label showing identification, calibration
date, next calibration date and the
calibrator’s stamp; apply any safeguard to
protect the instrument from adjustments that would
invalidate the calibration; return the instrument to
user with any
recommendations. |
The calibrated
instrument is tagged, logged, safeguarded from
adjustments, and returned |
| 10 |
Calibrations
Clerk |
When a
calibration date needs to be
extended |
Calibration log
or own judgment |
Calibration due
dates may be extended by no more than one-half of its
normal calibration period under circumstances deemed
appropriate; note the extension and circumstances in
calibration log. |
The calibration
date is extended for the calibrated instrument and
the calibration log
annotated |
| 11 |
Calibrations
Clerk |
When instruments
are not in use for extended periods or are
missing |
Personal
knowledge |
For an instrument
that is not in use it may be tagged as
“IA” (inactive), appropriately stored and
noted in the calibration log; tools that are missing
will be entered into the calibration log as such; a
“Tool Status” e-mail is to be sent to the
appropriate supervisor(s). |
All tools tagged
inactive or are missing are appropriately disposed.
The calibration log is updated and the affected
supervisor(s) are emailed |
| 12 |
Calibrations
Clerk, Engineer, or
Supervisor |
When an inactive
("IA") precision measuring instrument is
needed |
Personal
knowledge and the calibration tag on
instrument |
Go to step 4
unless the Calibrations Clerk is
not available; in the event that the Calibrations Clerk is
not available, and a tool is needed in an emergency,
a supervisor may temporarily activate the tool by
doing a calibration per step 6. An email must be sent
immediately to Calibrations
Clerk. When
the Calibrations Clerk becomes available, the tool will be submitted for
proper calibration and entry into the calibration
log. |
Calibrations Clerk calibrates and updates calibration control
system |
1 An instrument is
verified by taking measurements at the lower 25% AND
upper 75% (rounded to the nearest whole increment) of the
measuring instrument's measurement range. After taking
the above maximum measure, the measuring instrument must
be returned to zero (0) point and the reading
observed. All prescribed tolerances must be observed.
Deficiencies including tip wear that may cause the
instrument to measure incorrectly must be
monitored.
2
Includes, but is not restricted to any existing
calibrated and logged precision measuring device
(caliper, CMM, v-block, micrometer, scale, fixture,
gage etc), any precision measuring device not previously
entered into the calibration system, any previously
rejected precision measuring device that has been
repaired, or any individual required by TOPS (QC
Inspectors, Magnaflux inspectors, etc) to be entered
into the calibration system.
3 Calibrations
Clerk's calibrations are
per ANSI/NCSL Z540-1-1994. Readings must be taken at the
beginning point, mid point, end point and back to
beginning point of each instrument with no greater than
a three (3) inch increment between calibration points
unless otherwise specified by the QA Manager. On
calipers, tip-wear must be monitored by taking readings
well into the jaws as well as near the beginning of the
jaws. Tolerances for calipers and similar instruments
are +/- .003 inches in the Forging Department and +/-
.002 inches in the Machining Department. Tolerances for
micrometers are to +/- .002 inches in Forging
Department and +/-.0005 inches (with 10th graduations)
or +/- .001 inches (without 10th graduations) in the
Machine Department. Machine Department Gage Pins
verification is +/-.0003. Exceptions, other instruments
must be accurate to within 5% of the dimension
tolerance interval of the dimension being measured.
Exceptions must come from the QC
Supervisor.
4
Calibration frequency is 12 months for CMM, Surface
Tables, Master Blocks, CNC Machines, Thread Gages, and
1-5 years (based on usage) on Standard Blocks. Visual
Acuity examinations will be given within the first
ninety (90) days of hire date regardless of temporary
or full time employment status. This testing is
required for QC Inspectors, QC Leadman, Post Forge
Workers that work on the inspection table,or anyone
doing a visual inspection of a part as part of final
review. This testing will be administered annually and
the employee will be identified with an
“Eyeballs” sticker and calibration date on
their hard hat. Magnaflux Level II every 3
years. Calibration frequency on all other instruments is
6 months unless otherwise specified by the QA
Manager.
5Calibration Control
System includes but is not
restricted to Quality Control, the Calibration Log, the
Verification Log, and Gage
Manager.
6 Current Trinity
Forge calibration stickers include but are not
restricted to Trinity Forge calibration stickers and
third party calibration stickers bearing a Calibrations
Clerk signature or stamp.
7Flight Safety Requirements: All
measurement and test equipment used to inspect Flight
Safety Parts is discriminate to within ten percent
(10%) of the total tolerance spread for the feature
being inspected. For total tolerance spreads of less
than .001 inch, measurement equipment is discriminate
to within twenty percent (20%) of the spread. Equipment
less discriminate is used with the specific approval of
the customer.